Criticare 506N3:
Spot-check or continuously monitor three critical parameters; (SpO2, NIBP, and predictive temperature) on neonates to adults with a single monitor. Advanced technology—including CSI DOX™ or Nellcor OxiMax™ pulse oximetry, CSI ComfortCuff™ NIBP, and FILAC® FasTemp® predictive temperature—and exceptional ease of use make our 506N3 series monitors the perfect solution to your clinical challenges.
Designed to travel everywhere you do, the lightweight 506N3 monitor with integrated carry-handle attaches quickly and easily to IV pole mount or rolling stand. Available in eight configurations—with or without predictive temperature, with or without internal printer, with CSI DOX™ or Nellcor OxiMax pulse oximetry—the 506N3 series provides maximum versatility and cost-effectiveness.
Custom Configurations:
506DN3: DOX SpO2, NIBP, heart rate
506DNP3: DOX SpO2, NIBP, printer, heart rate
506DNT3: DOX SpO2, NIBP, temperature, heart rate
506DNTP3: DOX SpO2, NIBP, temperature, printer, heart rate
506LN3: Nellcor SpO2, NIBP, heart rate
506LNP3: Nellcor SpO2, NIBP, printer, heart rate
506LNT3: Nellcor SpO2, NIBP, temperature, heart rate
506LNTP3: Nellcor SpO2, NIBP, temperature, printer, heart rate
• Optional DOX or Nellcor OxiMax SpO2 technology.
• Proprietary ComfortCuff NIBP measures on inflation for faster, more comfortable readings.
• Optional FILAC FasTemp predictive temperature with predictive and monitoring modes delivers quick, accurate measurements.
• Audible and visual alarms provide instant notification.
• Lightweight design and versatile mounting options—adapts readily to department and transport requirements.
• Optional internal printer for demand, automatic interval, or trend reports.
• Smart battery management with rechargeable sealed lead acid battery
Tech Sheet:
http://www.csiusa.com/pdf/506N3.pdf
Non-Invasive Blood Pressure
Technique: Oscillometric measure upon inflation
Measurement Time: <40 seconds typical; 0 to 300mmHg;
standard adult cuff
Automatic Measurement
Cycles: 1 - 60 minute; 2 - 4 hour; STAT
Inflation Pressure Range: 30 to 300mmHg (Adult)
30 to 150mmHg (Pediatric)
20 to 150mmHg (Neonate)
Resolution: 1mmHg
Clinical Accuracy: SP-10:2002
Transducer Accuracy: ±2mmHg or ±2% of reading, whichever is
greater, over full range
STAT mode: 5 minutes of consecutive readings
Pulse Oximetry
Range: 1% to 99%
Resolution: 1%
Accuracy (DOX only): 70% to 99% range: ±2%;
50% to 69% range: ±3%;
<50%: unspecified
Accuracy (OxiMax Only): 70% to 99% range: ±2%;
<70%: unspecified
Method: Dual wavelength LED
Audio: Tone with each detected pulse; pitch varies
with saturation
Heart (Pulse) Rate
Range: 20 to 300bpm
Accuracy (DOX only): Greater of ±1bpm or ±1%, of reading
Unspecified for <30 or >240 bpm
Accuracy (OxiMax only): Greater of ±1bpm or ±1%, of reading
Unspecified for <30 or >240 bpm
Update Time: Every second
Resolution: 1bpm
Pulse Volume: 10 levels or Off
Temperature (optional)
Type: FILAC FasTemp
Method: Predictive or Continuous
Channels: 1
Range: 30.0º to 43.0º C
Accuracy: ±0.2º C for 30.0º to 36.9ºC;
±0.1º C for 37.0º to 39.0ºC;
±0.2º C for 39.1º to 43.0ºC
Resolution: 0.1º F or 0.1º C
Predictive Response Time: Oral: less than 10 seconds;
Axial: less than 10 seconds;
Rectal: less than 10 seconds
Probes/Wells: Oral/Axial FILAC FasTemp [202026]
Rectal FILAC FasTemp (optional) [202034]
Isolation Well, Oral/Axial [202089]
Isolation Well, Rectal (optional) [202087]